QPHY-ENET Operator’s Manual Rev G

1000Base-T Common Mode Output Voltage Test

This test ensures that the peak-to-peak common mode output voltage at the MDI is within conformance limits.
The peak-to-peak common mode output voltage is measured as the worst-case minimum to worst-case maximum
common mode output voltage. This test does not use a high-pass filter. This measurement is made for 4 pairs
(A,B,C,D) by changing J35, J36, J37, J38 jumper pins.

Figure 52. Fixture setup for Common Mode Output Voltage

1. Connect DUT to J67 and connect oscilloscope Channel 2 to J2.

2. Measure each Pair by changing Jumpers (A: J35, B: J36, C: J37, D: J38).

3. The transmitter signal must be less than 50 mV to pass the test.