background image

O

PERATOR

S

M

ANUAL

FlexRay-TDP-OM-E Rev A

31

Trigger Capability

Format

Hexadecimal or Binary for Frame ID
Decimal for Cycle Count

Trigger Setup


Trigger on TSS (Start), Frame ID, Cycle Count, Symbols and Errors

Symbols: Channel Idle Delimiter (CID) Symbol, Collision Avoidance Symbol (CAS), Media Access
Test Symbol (MTS), Wakeup Pattern (WUP)

Errors:
Frame Start Sequence (FSS) Error – triggers when the logic high time between the TSS and the
first byte is too long

Byte Start Sequence (BSS) Error – triggers anytime the BSS pattern is not seen between bytes
where expected

Frame End Sequence (FES) Error – triggers when the FS is not seen after the last byte

Frame ID and Cycle Count
Condition Setup

<=, <, =, >, >=, <>, in range, out of range, don’t care.

DATA Setup

Hexadecimal: # Data Bytes = 0 to 8. Data can be defined by nibble. Triggers on that data pattern
regardless of position or in user settable location.

Binary: Any combination of 0,1, or X for 1-64 bits. Triggers on that data pattern regardless of
position or in user settable location.

Bit Rates

2.5, 5.0 or 10 Mb/s selectable.

Trigger Input

Any analog Channel or the EXT input.

Trigger Design

Internal to oscilloscope, settable like any other oscilloscope trigger

Physical Layer Test

Eye Diagram

Eye diagram creation with mask testing at TP1 and TP4. Mask testing allows for 'Stop on Failure'

Measurements

Propagation Delay, Asymmetric Delay, Truncation, Jitter

Search Capability

Pattern Search

Search by Next ID, Next Frame, or Next Error in Hexadecimal formats

Other

Compatible With…

TDP (Trigger, Decode & Physical) and TD (Trigger & Decode) Option fully compatible with WRXi
(retrofit kits available and required for some WRXi oscilloscopes)

D (Decode) Option fully compatible with WR6000, WP7000, and WM8000 Series.