Performance verification, Required test equipment – Teledyne LeCroy ZS4000 User Manual

Page 21

background image

Operator’s Manual

923360-00 Rev A

15

Performance Verification

This procedure can be used to verify the warranted characteristics of the ZS4000
High Impedance Active Probe.

The recommended calibration interval for ZS series probes is one year. The
complete performance verification procedure should be performed as the first
step of annual calibration. Test results can be recorded on a photocopy of the
Test Record provided in Appendix A at the end of the manual.

Performance verification can be completed without removing the probe covers
or exposing the user to hazardous voltages. Adjustment should only be
attempted if a parameter measured in the Performance Verification Procedure is
outside the specification limits.

NOTE: Adjustment should only be performed by qualified personnel

This procedure tests the following specifications:

Output Zero Voltage

LF Attenuation Accuracy

Required Test Equipment

This procedure has been developed to minimize the number of calibrated test
instruments required. Because the input and output connector types may vary on
different brands and models of test instruments, additional adaptors or cables
may be required.

Description

Minimum Requirement

Test Equipment Examples

Digital Oscilloscope

ProBus Interface; 12-bit
vertical resolution

Teledyne LeCroy HDO6000
or HRO 6xZi series

Function Generator

Sine Wave output amplitude
adjustable to 14.14 Vp-p
(5 Vrms) into 1 MΩ at 70 Hz

Agilent Technologies 33120A
Stanford Research DS340

SMA Coaxial Cable

Male to Male, 50 Ω, 24" Cable

Pomona 4846-C-24
Pomona 4846-X-24
Pasternack PE3385LF-24

Adaptor SMA to BNC
(qty 2)

SMA (f) to BNC (m)

Pomona 4289
Pasternack PE9073

Calibration Fixture

50 Ω thru line with square pins

Teledyne LeCroy
PCF200