Performance verification, Required test equipment – Teledyne LeCroy ZS4000 User Manual
Page 21

Operator’s Manual
923360-00 Rev A
15
Performance Verification
This procedure can be used to verify the warranted characteristics of the ZS4000
High Impedance Active Probe.
The recommended calibration interval for ZS series probes is one year. The
complete performance verification procedure should be performed as the first
step of annual calibration. Test results can be recorded on a photocopy of the
Test Record provided in Appendix A at the end of the manual.
Performance verification can be completed without removing the probe covers
or exposing the user to hazardous voltages. Adjustment should only be
attempted if a parameter measured in the Performance Verification Procedure is
outside the specification limits.
NOTE: Adjustment should only be performed by qualified personnel
This procedure tests the following specifications:
Output Zero Voltage
LF Attenuation Accuracy
Required Test Equipment
This procedure has been developed to minimize the number of calibrated test
instruments required. Because the input and output connector types may vary on
different brands and models of test instruments, additional adaptors or cables
may be required.
Description
Minimum Requirement
Test Equipment Examples
Digital Oscilloscope
ProBus Interface; 12-bit
vertical resolution
Teledyne LeCroy HDO6000
or HRO 6xZi series
Function Generator
Sine Wave output amplitude
adjustable to 14.14 Vp-p
(5 Vrms) into 1 MΩ at 70 Hz
Agilent Technologies 33120A
Stanford Research DS340
SMA Coaxial Cable
Male to Male, 50 Ω, 24" Cable
Pomona 4846-C-24
Pomona 4846-X-24
Pasternack PE3385LF-24
Adaptor SMA to BNC
(qty 2)
SMA (f) to BNC (m)
Pomona 4289
Pasternack PE9073
Calibration Fixture
50 Ω thru line with square pins
Teledyne LeCroy
PCF200