Hfp1500 high frequency probe, Hfp1500 test record – Teledyne LeCroy HFP1500 User Manual
Page 46

HFP1500 High Frequency Probe
40
922252-00 Rev
A
HFP1500 TEST RECORD
Step
Description
Intermediate Data
Test Result
Output Zero Voltage
A-5
Output Zero Voltage (Test limit
≤± 800 μV)
__________ V
Offset Accuracy
B-7
Power Supply Negative Output Voltage
__________ V
B-10
Expected Negative Output Voltage
__________ V
B-14
Measured Negative Output Voltage
__________ V
B-16
Offset Error Voltage (Test limit
≤± 10.8 mV)
_________mV
B-21
Power Supply Positive Output Voltage
__________ V
B-23
Expected Positive Output Voltage
__________ V
B-26
Measured Positive Output Voltage
__________ V
B-28
Offset Error Voltage (Test limit
≤± 10.8 mV)
_________mV
LF Attenuation Accuracy
C-5
Generator Output Voltage
__________ V
C-6
Expected Output Voltage, top range
__________ V
C-8
Measured Output Voltage, top range
__________ V
C-10
Gain Error, top range (Test limit
≤± 1.0%)
__________%
C-15
Generator Output Voltage
__________ V
C-17
Expected Output Voltage, mid range
__________ V
C-19
Measured Output Voltage, mid range
__________ V
C-21
Gain Error, mid range (Test limit
≤± 1.0%)
__________%