Test Equipment Required
Table 3 List of Required Equipment lists the test equipment and accessories (or their equivalents)
that are required for complete calibration of the HFP1500 Probe. Specifications given for the test
equipment are the minimum necessary for accurate calibration. All test equipment is assumed to be
correctly calibrated and operating within the specification listed. Detailed operating instructions for
the test equipment are not given in this procedure. Refer to the test equipment manual if more
information is needed
If alternate test equipment is substituted, control settings or calibration equipment setups may
need to be altered.
Table 3 List of Required Equipment
Description Examples
Minimum Requirements
Example Test Equipment
Digital Oscilloscope
ProBus interface
Teledyne LeCroy WavePro960
Teledyne LeCroy LT344
Digital Multimeter (DMM)
with test probe leads
4.5 digit
DC: 0.1% Accuracy
AC: 0.1% Accuracy
Agilent Technologies 34401A
Fluke 8842A-09
Power Supply
0-12 V, settable to 10 mV
HP E3611A
BNC Coaxial Cable (2 ea.)
Male to Male, 50 Ω, 36" Cable
Pomona 2249-C-36
Pomona 5697-36
BNC Tee Connector
Male to Dual Female
Pomona 3285
Calibration Fixture
ProBus Extender Cable
Teledyne LeCroy PROBUS-
CF01
Terminator, Precision, BNC
50 Ω ± 0.05%
Teledyne LeCroy TERM-CF01
Banana Plug Adapter (2 ea.)
Female BNC to Dual Banana
Plug
Pomona 1269
BNC to Mini-grabber
BNC Male to Mini-grabber
Cable, 36"
Pomona 5187-C-36