WaveMaster 8 Zi/Zi-A Oscilloscopes

50

922133-00 Rev A

waits indefinitely for an unforthcoming segment. During that time,
no oscilloscope functions are accessible. By means of a timeout
value, however, the acquisition will be completed, the waveform
displayed, and control of the oscilloscope returned to the user after
the timeout has elapsed.

5. Under Display Settings, touch inside the Display mode control, and

select a sequence mode display from the pop-up menu.

6. Touch the S

INGLE

trigger front panel button.

Note: Once a single acquisition has started, you can interrupt the
acquisition at any time by pressing the S

INGLE

front panel button a

second time or by pressing the S

TOP

front panel button. In this case,

the segments already acquired will be retained in memory.

Zooming Segments in Sequence Mode

You can zoom individual segments easily using the Q

UICKZOOM

front panel

button. When you zoom, the zoom traces default to Segment 1. Channel
descriptors indicate the total number of segments acquired. Zoom
descriptors indicate [Seg #] and #Segments in the Zoom. You can scroll
through the segments using Z

OOM

front panel position knob.

Sample Zoom of Segments in Sequence Mode