Teledyne LeCroy PXD Series User Manual

Page 41

background image

901217-01

ISSUED: March 2003

37

Supported Drivers:

IVI-Scope Driver

LeCroy PXD Getting Started Application Program

ActiveX Control

LabView Driver

UPDATE RATE

Supports PCI Bus transfer rates up to 100 MB/s peak data rates.

GENERAL

Auto-Calibration: Ensures specified DC and timing accuracy.

Auto-Calibration Time: < 500 ms

Recommended Factory Calibration Interval: one year

Temperature

Operating: 0 to 40 °C when installed in a PXI chassis with a minimum airflow of 5 cfm (PXD 212, 512, 1021)
or 15 cfm (PXD 114, 214, 514, 522, 1022) provided to the air inlet of the Digitizer

Storage (Non-Op): -40 to +71°C

Humidity

Operating: 5 to 80% RH (non-condensing). Upper limit derates to 50% RH above 30 °C.

Storage (Non-op): 5 to 95% RH (non-condensing). Upper limit derates to 75% RH above 30 °C and 45% RH
above 40 °C.

Altitude

Operating: Up to 3,048 m (10,000 ft) at or below 25 °C

Storage (Non-op): Up to 12,192 m (40,000 ft)

Vibration

Operating: Random vibration, 0.31 g

rms

, 5 to 500 Hz, 15 minutes in each of 3 orthogonal axes

Non-operating: Random vibration, 2.4 g

rms

, 5 to 500 Hz, 15 minutes in each of 3 orthogonal axes

Functional Shock: 30 g

peak

, half sine, 11 ms, 3 shocks (positive and negative) in each of 3 orthogonal axes,

18 shocks total

Electromagnetic Compatibility: Conforms to EN 61326-1:1998 (Emissions and Immunity)

Safety: Conforms to EN 61010-1:2001 (Installation Category I, Pollution Degree 2)

Certifications: CE Approved