Teledyne LeCroy PXD Series User Manual
Page 41

901217-01
ISSUED: March 2003
37
Supported Drivers:
•
IVI-Scope Driver
•
LeCroy PXD Getting Started Application Program
•
ActiveX Control
•
LabView Driver
UPDATE RATE
Supports PCI Bus transfer rates up to 100 MB/s peak data rates.
GENERAL
Auto-Calibration: Ensures specified DC and timing accuracy.
Auto-Calibration Time: < 500 ms
Recommended Factory Calibration Interval: one year
Temperature
Operating: 0 to 40 °C when installed in a PXI chassis with a minimum airflow of 5 cfm (PXD 212, 512, 1021)
or 15 cfm (PXD 114, 214, 514, 522, 1022) provided to the air inlet of the Digitizer
Storage (Non-Op): -40 to +71°C
Humidity
Operating: 5 to 80% RH (non-condensing). Upper limit derates to 50% RH above 30 °C.
Storage (Non-op): 5 to 95% RH (non-condensing). Upper limit derates to 75% RH above 30 °C and 45% RH
above 40 °C.
Altitude
Operating: Up to 3,048 m (10,000 ft) at or below 25 °C
Storage (Non-op): Up to 12,192 m (40,000 ft)
Vibration
Operating: Random vibration, 0.31 g
rms
, 5 to 500 Hz, 15 minutes in each of 3 orthogonal axes
Non-operating: Random vibration, 2.4 g
rms
, 5 to 500 Hz, 15 minutes in each of 3 orthogonal axes
Functional Shock: 30 g
peak
, half sine, 11 ms, 3 shocks (positive and negative) in each of 3 orthogonal axes,
18 shocks total
Electromagnetic Compatibility: Conforms to EN 61326-1:1998 (Emissions and Immunity)
Safety: Conforms to EN 61010-1:2001 (Installation Category I, Pollution Degree 2)
Certifications: CE Approved