Teledyne LeCroy PXD Series User Manual
Page 39

901217-01
ISSUED: March 2003
35
Offset Range:
±1 V (40 mV to 792 mV FSR, 50 ohms only)
±10 V (800 mV to 8 V FSR, 50 ohms only)
±1 V (40 mV to 800 mV FSR, 1 Mohms only)
±20 V (816 mV to 8 V FSR, 1 Mohms only)
Variable Gain Range: 0.1 to 1.0 of full scale
Variable Gain Resolution: 0.1% of full scale
Input Coupling:
Z
in
= 50 ohms DC, GND
Z
in
= 1 Mohms AC, DC, GND
AC Coupled Lower Cutoff:
< 10 Hz, frequency -3 dB
Input Impedance:
Z
in
= 50 ohms ±1.5%
Z
in
= 1 Mohms ±1.5% || 16 ± 2 pF
DC Accuracy: ±(2% full scale + 1.6% offset setting + 1 mV) @ gain >= 80 mV FSR
Vertical Resolution: 8 bits
Maximum Input Voltage:
Z
in
= 50 ohms, 5 V
rms
(including DC)
Z
in
= 1 Mohms, 100 V (DC + pk AC, frequency <= 5 kHz)
Input Connector(s): BNC (grounded)
ACQUISITION MODES
Single Shot: For transient and repetitive signals: 1 GS/s for models 114, 214, 514, 212, 512; 2 GS/s for
models 1021, 1022, 522
Sampling period settable in 1-2-4 sequence.(e.g., PXD512 period: 1 ns/pt, 2 ns/pt, 4 ns/pt, etc.)
Random Interleaved Sampling (RIS): For repetitive signals: up to 50 GS/s
Sequence: Stores multiple events, each of them time stamped (1 ns resolution) in segmented acquisition
memory.
Minimum Segment Length: 256 samples
Maximum Segment Length: 1 million samples
TIMEBASE SYSTEM
Capture Window at Maximum Sample Rate: up to 4 ms
Clock Accuracy: 10 ppm