Specifications – SKC Limited AirChek 3000 Deluxe Pump Operating Instructions User Manual
Page 9
www.skcltd.com
210-3311M Issue E
Page 7
Specifications
• Operating humidity: 0 to 95 % RH non-condensing
• Operating altitude: 2285 metres (7500 ft) maximum
• Timing accuracy: 1 minute per month at 25 °C
• Atmospheric pressure accuracy: ±10 mbar (± 0.3 “ H
g
)
• Temperature accuracy: ±5 % of reading
• Typical run times:
Back Pressure
Run Time at Stated Flow Rate (hours)
kPa
“ H
2
O
1.0 l/min
1.5 l/min
2.0 l/min
2.5 l/min
3.0 l/min
3.25 l/min
1.2
5
28
24
20
18
16
15
2.5
10
19
17
14
14
13
12
3.7
15
15
14
12
11
11
10
5.0
20
12
11
10
10
9
-
6.2
25
10
10
9
8
-
-
7.5
30
9
9
8
-
-
-
8.7
35
8
-
-
-
-
-
Note: Run times quoted are based on a new, fully charged battery, do not take account of increasing back