Care of the Battery Pack
Battery Testing
1. Connect SKC approved charger to battery pack. If charger indicator LED illuminates, battery pack charger
input is ok. If charger LED does not illuminate, battery pack input fuse is blown - Replace battery pack.
2. Leave battery pack connected to charger to fully charge.
3. If pump does not function at all after full charge of battery pack, battery pack output fuse has blown, or
battery cells have failed or are at end of life - Replace battery pack.
4. If pump functions after full charge of battery pack but gives significantly reduced run times before low
battery shutdown, battery cells are failing or are at end of life - Replace battery pack.
Battery Replacement
Refer to the pump component diagram on page 46. Unfasten the two security screws securing the battery
pack to the pump using the supplied 2mm A/F Allen key. Carefully pull the battery pack downwards to
disconnect from the pump case.
Fit the replacement battery pack to the pump, taking care to ensure that the battery output connector engages
with the socket in the base of the pump case. Secure the battery pack with the two security screws. Charge
the new battery pack fully before use.
Battery Disposal
• The EC Battery Directive and equivalent legislation in other countries requires that all batteries and battery
packs are disposed of correctly at the end of their working life. This means that they must be collected and
treated separately from other waste.
• Please ensure that any end-of-life SKC battery packs are collected and recycled or disposed of correctly.