CalChek Calibration Error Code Troubleshooting
Error Code Problem
Corrective Action
E 46
Full calibration - data analysis error.
Repeat the full calibration.
If this does not resolve the problem contact
SKC Ltd customer services for assistance.
E 47
Full calibration - Battery charge level is too
low. Less than two bars displayed on the
battery status indicator.
Fully charge the battery before repeating the
full calibration.
E 48
Single point calibration - Pump could not
achieve a successful calibration within 5
Defender flow readings.
Repeat the calibration.
If the problem repeats perform a full calibration.