Specifications
Pump Features
Flow fault function
Indicates flow fault due to obstructed tubing or excessive filter loading. Shuts
the pump down and enters ‘HOLD’ mode if the condition persists for longer
than 15 seconds. Automatically attempts to restart every five minutes (up to
a maximum of 10 restarts), until the flow fault condition clears in which case
normal running will resume.
Programmable run time
Sample run time programmable in minutes, via keypad and LCD screen.
Pump automatically shuts down at end of sample and retains run time data.
Programmable delayed
start time
Sample start delay time programmable in hours and minutes, via keypad and
LCD screen.
Particulate trap
Built in replaceable filter to trap particles that would otherwise contaminate
the pump mechanism.
Automatic air flow rate
calibration
With optional Defender primary calibrator and CalChek Communicator cable.
Enables automatic calibration of the sample air flow rate to the required level.
PC Connectivity
With optional DataTrac USB adapter cable and PC software. Enables
programming of timed sample runs, delayed starts and intermittent sampling.
Enables retrieval of pump run-time data and history to the PC.