Run Time Data Screens
When in HOLD or RUN modes press the button to scroll through the available run-time data screens:
Elapsed run time since last reset in minutes up to 9999 minutes (approximately 1
week). Above 9999 minutes display reverts to 0 minutes and continues to count.
Flow rate in millilitres/minute.
Accumulated sample air volume since last reset in millilitres or litres up to 9999 litres.
Above 9999 litres display reverts to 0 litres and continues to count.
Real time sample air temperature in °C or °F.
Clock time in 12 or 24 hour format with AM and PM indicators.
Real time atmospheric pressure in mbar, mm H
g
or “ H
g
.