Operation PEM-9004

Teledyne Analytical Instruments

44

CO concentration is below a factory preset threshold the value of which
depends on the sensor installed.

Whenever the CO pump is on (whether it is started automatically or

manually) a scored out CO-symbol appears in the status line as shown in
the screen display below.

.

If the CO-flushing-pump starts automatically due to an excess

concentration of CO it cannot be switched off manually until the CO-
concentration falls below a factory preset value.

Once the CO-concentration has reached the lower range the CO-

flushing-pump will automatically shut off.

If the analyzer is equipped with two CO-sensors the result of the

higher range sensor will be displayed when the lower range sensor is
flushed.

The active CO-flushing-pump has no influence on any other

sensors within the analyzer.

4.5.1.14

P

RINT

The measured data can be printed out by means of a wireless

infrared printer.

The printer in use can be selected from the Configuration menu.

See Section 4.4.3.3. The print speed depends on the printer selected.
Make sure the proper printer is activated to avoid possible failures while
printing.