Teledyne 5000B - Near Infrared Photometer User Manual

Page 87

background image

4–19

Operations/Analysis Unit 4

Teledyne Analytical Instruments

A Business Unit of Teledyne Electronic Technologies

Figure 4-1.

Optical System

4.4.2 Sample cell

The sample cell, generally constructed of 316SS, is located in the path of

the NIR radiation, between the source and the detector modules. Each com-
pound in the sample path exhibits its own characteristic absorption spectrum.
Cell spacer thicknesses will vary depending upon the absorbance of the compo-
nent of interest at the measuring wavelength. Due to the possible variation of
absorption with temperature, it is either necessary to maintain the sample at a
constant temperature during analysis or to compensate for the minor frequency
shifts of the water bonding which can change due to sample temperature varia-
tions.. To achieve this, two separate methods of temperature control are em-
ployed for constant temperature control.

1. A preheater (heat exchanger coil) is used on the incoming sample

stream to raise or lower it to the desired temperature controlled
level.