beautypg.com

Teledyne 5000B - Near Infrared Photometer User Manual

Page 45

background image

Photometric Analyzer

Operation /Control Unit 3

Part I 3-17

Teledyne Analytical Instruments

zero calibration is to be performed automatically or manually. Use the DÑ
arrow keys to toggle between AUTO and MAN zero settling. Stop when
AUTO appears, blinking, on the display.

Select zero
mode: AUTO

Press Enter to begin zeroing.

####.## %

SO2

Slope=#.### C—Zero

The beginning zero level is shown in the upper left corner of the dis-

play. As the zero reading settles, the screen displays and updates information
on Slope= in percent/second (unless the Slope starts within the acceptable
zero range and does not need to settle further). The system first does a coarse
zero, shown in the lower right corner of the screen as C—Zero, for 3 min, and
then does a fine zero, and displays F—Zero, for 3 min.

Then, and whenever Slope is less than 0.01 for at least 3 min, instead of

Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to Analyze. Software zero is indicated by S–
Zero in the lower right corner.

####.## % SO2
4 Left=#.### S—Zero

The zeroing process will automatically conclude when the output is

within the acceptable range for a good zero. Then the analyzer automatically
returns to the Analyze mode.

3.4.1.2

Manual Mode Zeroing

Press Zero to enter the Zero function. The screen that appears allows

you to select between automatic or manual zero calibration. Use the DÑ
keys to toggle between AUTO and MAN zero settling. Stop when MANU-
AL appears, blinking, on the display.

Select zero
mode: MANUAL

Press Enter to begin the zero calibration. After a few seconds the first

of three zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate.