Control unit, 8 part i – Teledyne 5000B - Near Infrared Photometer User Manual
Page 36
3 Operation
/Control Unit
Model 5000B
3-8 Part I
Teledyne Analytical Instruments
When you have finished typing the new password, press Enter. A verifica-
tion screen appears. The screen will prompt you to retype your password for
verification.
Enter PWD To Verify:
A A A
Use the arrow keys to retype your password and press Enter when
finished. Your password will be stored in the microprocessor and the system
will immediately switch to the Analyze screen, and you now have access to
all instrument functions.
If all alarms are defeated, the Analyze screen appears as:
1.95 ppm SO
2
nR1: Ш — 1Ш Anlz
If an alarm is tripped, the second line will change to show which alarm
it is:
1.95 ppm SO
2
AL—1
NOTE:
If you log off the system using the LOGOUT function in the
system menu, you will now be required to re-enter the password to
gain access to Alarm, and Range functions.
3.3.3
Logging Out
The LOGOUT function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the instrument
off. By entering LOGOUT, you effectively log off the instrument leaving
the system protected against use until the password is reentered. To log out,
press the System button to enter the System function.
DIG_FILT SELF-TEST
PWD LOGOUT MORE
Use the < > arrow keys to position the blinking over the LOGOUT
function, and press Enter to Log out. The screen will display the message:
Protected until
password entered