Photometric Analyzer

Operation /Control Unit 3

Part I 3-5

Teledyne Analytical Instruments

Note: Before setting up an AUTO-CAL, be sure you understand the Zero

and Span functions as described in section 4.4, and follow the
precautions given there.

Note: If you require highly accurate AUTO-CAL timing, use external

AUTO-CAL control where possible. The internal clock in the
Model 5000BF is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.

To setup an Auto–Cal cycle:

Choose System from the Function buttons. TheVFD will display five

subfunctions.

DIG_FILT SELF—TEST
PWD LOGOUT MORE

Select MORE and press the Enter Key

AUTOCAL

FILSOL

HOLD

CAL-HOLD-TIMER

MORE

Use < > arrows to blink

AUTO—CAL

, and press Enter. A new screen

for ZERO/SPAN set appears.

ZERO in Ød Øh off
SPAN in Ød Øh off

Press < > arrows to blink

ZERO

(or

SPAN

), then press Enter again.

(You won’t be able to set OFF to ON if a zero interval is entered.) A Span
Every ... (or Zero Every ...) screen appears.

Zero schedule: OFF
Day: Ød Hour: Øh

Use DÑ arrows to set an interval value, then use < > arrows to move to

the start-time value. Use DÑ arrows to set a start-time value.

To turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL):

Press System again, choose AUTO–CAL

,

and press Enter again. When the

ZERO/SPAN values screen appears, use the < > arrows to blink the ZERO
(or SPAN) and press Enter to go to the next screen. Use < > to select OFF/
ON field. Use DÑ arrows to set the OFF/ON field to ON. You can now turn
these fields ON because there is a nonzero span interval defined.

If instrument is turned off, the next time the instrument is powered, the

instrument will automatically perform a calibration cycle after 3 minutes of
entering the sample mode if AUTOCAL functions were on prior to shut
down.