1 Introduction

Model 5000B

Normally open

Moving contact

Normally closed
Normally open
Moving contact

Figure

2-6: Types of Relay Contacts

The connectors are:

Threshold Alarm 1:

• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).

Threshold Alarm 2:

• Can be configured as high (actuates when concen-

tration is above threshold), or low (actuates when
concentration is below threshold).

• Can be configured as fail-safe or non-fail-safe.

• Can be configured as latching or nonlatching.
• Can be configured out (defeated).

System Alarm:

Actuates when DC power supplied to circuits is

unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be de-
feated. Actuates if self test fails.

To reset a System Alarm during installation, discon-

nect power to the instrument and then reconnect it

Further detail can be found in chapter 3, section 3-5.

Digital Remote Cal Inputs

Remote Zero and Span Inputs: The REMOTE SPAN and RE-

MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They