2-4: Part I

Teledyne Analytical Instruments

1 Introduction

Model 5000B

4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =

Range 3.

Figure 2-5: Analog Output Connections

Examples:

The analog output signal has a voltage which depends on the sample

concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.

The signaloutput for concentration is linear over currently selected

analysis range. For example, if the analyzer is set on a range that was
defined as 0-10 %, then the output would be as shown in Table 2-1.

Table 2-1: Analog Concentration Output-Examples

Concentration

Voltage Signal

Current Signal

%

Output (V dc)

Output (mA dc)

0

0.0

4.0

1

0.1

5.6

2

0.2

7.2