beautypg.com

Teledyne 5000B - Near Infrared Photometer User Manual

Page 16

background image

1-8: Part I

Teledyne Analytical Instruments

1 Introduction

Model 5000B

Alphanumeric Interface Screen: The backlit VFD screen is an easy-

to-use interface between operator and analyzer. It displays values, options,
and messages for immediate feedback to the operator.

1.5

Control Section Interface Panel

The Control Section interface panel, shown in Figure 1-2, contains the

electrical terminal blocks for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter of
this manual.

Power Connection

AC power source, 115VAC, 50/60 Hz

Analog Outputs

0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.

Alarm Connections

2 concentration alarms and 1 system
alarm.

RS-232 Port

Serial digital concentration signal output
and control input.

Remote Bench

Provides all electrical interconnect to the
Analysis Section.

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration.

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

Range ID Contacts

Four separate, dedicated, range relay
contacts.

Network I/O

Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 5000B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.