beautypg.com

Teledyne 5000B - Near Infrared Photometer User Manual

Page 28

background image

2-10: Part I

Teledyne Analytical Instruments

1 Introduction

Model 5000B

Check that sample pressure typically between 0 and 30 psig,
according to the requirements of your process.

Power up the system, and test it by performing the following operation:

1. Repeat the Self-Diagnostic Test.

2. Zero the instrument.

3. Span the instrument.

Consult calibration recommendation in the addendum for your

particular application whether liquid or gas phase monitoring.

For steps 2 and 3, refer to part II for calibration.