Photometric Analyzer

Operation /Control Unit 3

Part I 3-19

Teledyne Analytical Instruments

3.4.2

Span Cal

The Span function on the menu is used to span calibrate the analyzer. Span

calibration can be performed in either the automatic or manual mode.

Make sure the span fluid is flowing to the instrument.

3.4.2.1

Auto Mode Spanning

Observe all precautions in sections 3.4 and 3.4.2, above. Press Span to

enter the span function. The screen that appears allows you to select whether
the span calibration is to be performed automatically or manually. Use the
DÑ arrow keys to toggle between AUTO and MAN span settling. Stop
when AUTO appears, blinking, on the display.

Select span
mode: AUTO

Press Enter to move to the next screen.

Span Val: 2Ш.ШШ %
To begin span

Use the < > arrow keys to toggle between the span concentration value

and the units field (%/ppm). Use the DÑ arrow keys change the value and/
or the units, as necessary. When you have set the concentration of the span
fluid you are using, press Enter to begin the Span calibration.

####.##%

SO2

Slope=#.### Span

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates informa-
tion on Slope. Spanning automatically ends when the span output corre-
sponds, within tolerance, to the value of the span fluid concentration. Then
the instrument automatically returns to the analyze mode.

3.4.2.2

Manual Mode Spanning

Press Span to start the Span function. The screen that appears allows

you to select whether the span calibration is to be performed automatically or
manually.

Select span
mode: MANUAL