Teledyne 4020 – Hydrocarbon analyzer (special version for automatic background gas switching) User Manual

Page 98

background image

Appendix Model

4020

Teledyne

Analytical

Instruments

80

Application Data

Model

Number:

Serial

Number:

Ranges

Of

Analysis:

Output

Signal:

Start-Up Span Setting:

Control Panel Pressure Settings:

Comments:

Blanket Air:

Fuel:

Sample:

Bypass Flowrate: A minimum of 0.25 SCFH is indicated on

the integral flowmeter

Recommended Support Gases: The accessory gases should all be

laboratory analyzed and certified as to
their hydrocarbon content.

Blanket Air: Water pumped, hydrocarbon free,

compressed air.

Zero Gas: Water pumped, hydrocarbon free

(certified)

Span Gas: ppm methane in

(certified)

The span gas contains a known amount of methane impurity. The

span gas must be certified as to its methane equivalent impurity.

IMPORTANT: For the best possible results, zero and span

gases that are representative of the parent (or
background) gas should always be used.

This is particularly true in applications where the analyzer is being used

to analyze a number of different samples with varying concentrations of
oxygen. The geometry and temperature of the flame will be somewhat
dependent on the oxygen content of the sample. Alterations in the flame
configuration will lead to differences in output readings. For the best possible
results, TAI recommends that zero and span gases representative of each of
these samples be used to standardize the analyzer.