Total Hydrocarbon Analyzer

Maintenance

Teledyne

Analytical

Instruments

77

standoffs, instead of directly to the printed circuit, to eliminate the
possibility of leakage currents.

To eliminate any possibility of contamination of the insulating

materials employed, the completed PC board is ultrasonically cleaned in
laboratory grade alcohol. Under no circumstances should the parts
described be handled with bare fingers
. A freshly-scrubbed finger,
stroked along one of the glass resistors, would deposit enough skin oil to
completely upset the range division of the attenuator circuit.

Resistor R3 is a 1000 MΩ resistor used in the feedback circuit of the

amplifier. R2 has a resistance of 10,000 MW and is used in series with
the zero potentiometer slider. This circuit is used to nullify any offset
signal introduced by the signal electrode. Trimmer P1 is used to nullify
the offset signal generated by the electrometer amplifier.

The output of the circuit is standardized against gases with known

hydrocarbon concentrations by zero and span calibration, so that the
meter and/or recorder indicates the hydrocarbon concentration of the gas
being used.

The positive and negative operating voltage required by the

electrometer amplifier is furnished by a switching power supply
circuit,mounting at the back panel of the analyzer

The stability of the electrometer circuit can be tested as follows:

1. Disconnect the collector cable.

2. Place the analyzer in the auto range.

3. Adjust zero offset value so that the recorder reads at some

point upscale, and record a 24 hour chart.