Total Hydrocarbon Analyzer

Installation

Teledyne

Analytical

Instruments

21

background gas.
Relay is ENERGIZED when the background gas
selection is O2, DE-ENERGIZED when the
background gas selection is N2.

Note: The polarity of this relay can be reversed using the A2

Match/A2 Invert function available from the Background
Gas display menu. See Section 4.7.2.3.

Table 3-4: Alarm Relay Contact Pins

Pin Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Background gas selection, (normally closed contact)

44

Background gas, moving contact

43

Background gas selection, (normally open contact)

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact

3.3.3.6

B

ACKGROUND

G

AS

S

ELECTION AND

D

IGITAL

R

EMOTE

C

AL

I

NPUTS

Background Gas Selection

When Background Gas is set to Auto (see Section 4.7.2), the 0-24 V

input formerly used for the Remote Zero command will be used to
select the background gas as follows:

Voltage Gas

Selected

0V N

2

selected

24V O

2

selected

The background gas selection inputs (formerly digital remote zero

input) accepts 0 V (off) or 24 VDC (on) for automatic control of the
background gas switching. The customer will provide a signal that is