beautypg.com

Teledyne 4020 – Hydrocarbon analyzer (special version for automatic background gas switching) User Manual

Page 82

background image

Operation Model

4020

Teledyne

Analytical

Instruments

64

performed prior to using selected sample/background gas
for analysis.

 Auto: Automatically select the sample/background gas that is

used by the analyzer during calibration or measurement. In
Auto mode, the sample/background gas will be determined
by the signal received at the remote zero contacts as
described in Section 3.3.3.6. The AL-2 relay contact will
also switch in conjunction with the signal input to the remote
zero contacts.

Note: It is the user’s responsibility to ensure proper valves are

installed for automatically switching the proper
sample/background gas to the analyzer synchronized to
the signals to and from the 4020 Analyzer. See Section
3.3.3.6.

 Match/Invert: This function reverses the logic of the AL-2

relay contact (formerly alarm 2) and is used as an indication
of the current selected background gas.

4.7.2.1

M

ANUALLY

S

ELECTING A

B

ACKGROUND

G

AS

To manually select a background gas:

<▲▼>

From the MAIN MENU, scroll to the BACKGROUND
GAS function using the UP/DOWN keys.

<Enter>

Pressing activates the function and takes you to
the next screen.

<Enter>

Use the key to move over to the gas field (O2,

N2, or Auto).

<▲▼>

Use the UP/DOWN keys to toggle between the options..

<Enter>

Use the key to accept the appropriate
background gas when it is displayed. Note that the AL-2
contacts will switch accordingly. See Section 4.7.2.3
Match/Invert.