Total Hydrocarbon Analyzer

Operation

Teledyne

Analytical

Instruments

61

4.6.16.4

O

BSERVING THE GAS

S

TREAM

C

ONCENTRATIONS

The actual gas concentrations of the gas streams can be observed

using two different methods:

1. Use the USER SELECT option of the VALVE

SELECTIONS menu and choose the appropriate valve to
open from the second line of the display. Then return to
Analysis mode by repeatedly pressing and observe
the concentration of that stream on the display.

Note: With this technique, the alarms are not suppressed and

depending on the gas and the alarm setpoint, it could
trigger an alarm.

2. Use the DEFAULT option of the VALVE SELECTIONS

menu and select the SPAN (or ZERO) function from the
MAIN MENU. With the SPAN (or ZERO) function
selected, select MANUAL FINISH from the first line of the
display on the SPAN or ZERO submenu. The display will
show:


Or

Note: The actual displayed concentration may be different

depending on your calibration gas.

Then begin the span or zero routine and observe the
concentration on the screen. Since the analyzer is in a calibration
mode, the alarms will be suppressed.