Total Hydrocarbon Analyzer

Operation

Teledyne

Analytical

Instruments

31

2. Allow at least 2 hours warm-up (heat up sensor & sample

system) after making the air adjustment described below. Warm
up time is set by the software at the factory.

DO NOT attempt to ignite the flame during warm up
countdown
. Condensation will occur.

4.4 Activating the Support Gases

4.4.1 Air

1. Set the air tank regulator to 30 psig.

2. Adjust the instrument air regulator until the air pressure

gauge reads the recommended air pressure.

After the air is flowing through the sensor and warm-up time has

been completed, activate the following gases:

4.4.2 Sample Gas

Set the sample gas tank regulator to 30 psig (or a pressure which
matches the sample pressure) and adjust the instrument sample
regulator until its (sample) pressure gauge reads the recommended
sample pressure.

4.4.3 Span Gas

1. Feed span gas to the analyzer. This instrument has an automatic

valve selection feature for directing sample or calibration gas to
the analyzer. See Section 4.6.16 for setup and operation of the
valve selection feature..

2. Set the span gas regulator to 30 psig or to match the sample

pressure.

3. Observe that the instrument sample pressure gauge still reads the

recommended sample pressure and that the bypass flowmeter
reads from 0.5 to 1.0 SCFH.

4.4.4 Fuel

1. Open the main valve on the tank and set the fuel tank pressure

regulator to 30 psig (or to match sample pressure).