Main Features of the Analyzer
The Model 3010TAC series Oxygen Analyzers are sophisticated yet
simple to use. The main features of these analyzers include:
•
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•
High resolution, accurate readings of oxygen content: from low
ppm levels through 25%. Large, bright, meter readout.
•
Stainless steel cell block is standard equipment.
•
Advance design Micro-Fuel Cell sensor, Specially designed
model for trace analysis, has a one year warranty and an
expected lifetime of two years.
•
Versatile analysis over a wide range of applications.
•
Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•
Three user definable output ranges allow best match to users
process and equipment: 0-10 ppm through 0-250,000 ppm for
trace analysis.
•
Air-calibration range for convenient spanning at 20.9 %.
•
Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•
Two adjustable concentration alarms and a system failure alarm.
•
Self-diagnostic testing, at startup and on demand, with continuous
power-supply monitoring.
•
CE Compliance (EMC & LVD).
•
RS-232 serial digital port for use with a computer or other digital
communications device.
•
Analog outputs for Concentration and Analysis Range: 0–1 V dc
standard. Isolated 4–20 mA dc standard.
•
Compact and versatile design: flush-panel, rack-mountable, or
bulkhead mounted Control Units available.