Oxygen Analyzer

Teledyne Analytical Instruments

Preface

Overview

The Teledyne Analytical Instruments Model 3010TAC Trace Oxygen

Analyzer is a versatile microprocessor-based instrument for detecting parts-
per-million (ppm) levels of oxygen in a variety of background gases. It is a
“split architecture” instrument. This means that a general purpose Control
Unit, designed for nonhazardous areas only, remotely controls a specially
designed Analysis Unit, or remote probe, that can operate in a hazardous
area. 3010TAC Analyzers meet or exceed all of the requirements of the
Commonwealth of Europe (CE) for Radio Frequency Interference and
Electromagnetic Interference (RFI/EMI) protection, and Low Voltage
Directive (LVD).

Part I of this manual covers the Model 3010TAC General Purpose flush-

panel and/or rack-mount Control Unit only. This Control Unit is for indoor
use in a nonhazardous environment. The Intrinsic Safe Barriers assy is de-
signed for indoor/outdoor nonhazardous environments only. The Analysis
Units (or Remote Probes), is for hazardous environments. Part II of this
manual covers the 3010TAC Remote Probe.

Typical Applications

A few typical applications of the Model 3010TAC are:

Monitoring inert gas blanketing

Air separation and liquefaction

Chemical reaction monitoring

Semiconductor manufacturing

Petrochemical process control

Quality assurance

Gas analysis certification.