Teledyne 3010TAC - Split architecture trace oxygen analyzer (EU - CENELEC) User Manual
Page 15
Part I: 1-3
Oxygen Analyzer
Part I: Control Unit
Teledyne Analytical Instruments
•
Enter
Moves VFD display on to the next screen in a series.
If none remains, returns to the
Analyze
screen.
•
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the
Analyze
screen.
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any lighting.
It is accurate across all analysis ranges. The –TA models produce
continuous readout from 0-10,000 ppm and then switch to continuous
percent readout from 1-25 %.
Alphanumeric Interface Screen: The backlit VFD screen is an easy-
to-use interface between operator and analyzer. It displays values, options,
and messages that give the operator immediate feedback.
Standby Button: The Standby turns off the display and outputs,
but circuitry is still operating.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits .
1.3 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
1.4
Control Unit Rear Panel
The Control Unit rear panel, shown in Figure 1-2, contains the
electrical connectors for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter
of this manual.