Oxygen Analyzer

Part I: Control Unit

Part I: 3-5

Teledyne Analytical Instruments

Analog Outputs: There are four DC output signal pins—two pins

per output. For polarity, see Table 3-1. The outputs are:

0–1 V dc % of Range: Voltage rises linearly with increasing oxygen,

from 0 V at 0 ppm to 1 V at full scale ppm. (Full
scale = 100% of programmable range.)

0–1 V dc Range ID:

0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.

4–20 mA dc % Range: Current increases linearly with increasing oxygen,

from 4 mA at 0 ppm to 20 mA at full scale ppm.
(Full scale = 100% of programmable range.)

4–20 mA dc Range ID: 8 mA = Low Range, 12 mA = Medium Range, 16

mA = High Range, 20 mA = Air Cal Range.

Table 3-1: Analog Output Connections

Pin

Function

3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ % Range, 4-20 mA, floating

6

– % Range, 4-20 mA, floating

8

+ Range ID, 0-1 V dc

23

– Range ID, 0-1 V dc, negative ground

24

+ % Range, 0-1 V dc

7

– % Range, 0-1 V dc, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the

internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 V ac into a
resistive load. The connectors are:

Threshold Alarm 1:

• Can be configured as high (actuates when concen-

tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as failsafe or nonfailsafe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).

Threshold Alarm 2:

• Can be configured as high (actuates when concen-

tration is above threshold), or low (actuates when
concentration is below threshold).