beautypg.com

Teledyne 3010TAC - Split architecture trace oxygen analyzer (EU - CENELEC) User Manual

Page 32

background image

3 Installation

Model 3010TAC

3-10: Part I

Teledyne Analytical Instruments

Table 3-5: Commands via RS-232 Input

Command

Description

as

Immediately starts an autospan.

az

Immediately starts an autozero.

st

Toggling input. Stops/Starts any status message output
from the RS-232, until st is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-6 lists certain RS-232 values that are required by the 3010TAC
implementation.

Table 3-6: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

3.3.4 Remote Probe Connection

The Models 3010TAC are split architecture (dual-chassis) instru-

ments, which have a Remote Probe, or Analysis Unit. The remote probe is
for receiving the oxygen sensor and thermistor signals. See Figure 3-7 for
remote probe connection. The connections of the Analysis Unit are covered
in detail in Part II, section 3.4, of this manual.

Remote Probe 9-pin Connector

Figure 3-7: Remote Probe Interface Pinout

8

5

6

1

Sensor Hot

Sensor Return

Thermistor Return

Thermistor Hot