Teledyne 3010TAC - Split architecture trace oxygen analyzer (EU - CENELEC) User Manual
Page 55
Oxygen Analyzer
Part I: Control Unit
Part I: 4-21
Teledyne Analytical Instruments
concentration and one range ID) and two isolated 4-20 mA dc current
outputs (one concentration and one range ID).
See Rear Panel in chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on range that was defined
as 0–100 ppm O
2
, then the output would be:
Voltage Signal
Current Signal
ppm O
2
Output (V dc)
Output (mA dc)
0
0.0
4.0
10
0.1
5.6
20
0.2
7.2
30
0.3
8.8
40
0.4
10.4
50
0.5
12.0
60
0.6
13.6
70
0.7
15.2
80
0.8
16.8
90
0.9
18.4
100
1.0
20.0
The analog output signal has a voltage which depends on the oxygen
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. The following
table gives the range ID output for each analysis range:
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
12
HI
0.75
16
CAL (0-25%)
1.00
20