Teledyne 3010TAC - Split architecture trace oxygen analyzer (EU - CENELEC) User Manual
Page 29
Oxygen Analyzer
Part I: Control Unit
Part I: 3-7
Teledyne Analytical Instruments
synchronous signal must open and close external span valve
appropriately. See Figure 3-5 Remote Probe Connector.
(The –C option internal valves operate automatically.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Table 3-3: Remote Calibration Connections
Pin
Function
9
+ Remote Zero
11
– Remote Zero
10
+ Remote Span
12
– Remote Span
40
Cal Contact
41
Cal Contact
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3010TAC Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span
is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Valve connections (described below) provides