Oxygen Analyzer

Part II: Analysis Unit

Part II: 4-3

Teledyne Analytical Instruments

If a cell was working satisfactorily, but ceases to function before the

warranty period expires, the customer will receive credit toward the purchase
of a new cell.

If you have a warranty claim, you must return the cell in question to the

factory for evaluation. If it is determined that failure is due to faulty work-
manship or material, the cell will be replaced at no cost to you.

Note: Evidence of damage due to tampering or mishandling will

render the cell warranty null and void.

4.4

System Self Diagnostic Test

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to More, and press

Enter

.

3. Use the < > arrow keys to move to Self-Test, and press

Enter

.

4. Observe the error-code readings on the VFD Display screen, and

check Table 5-1, below, to interpret the codes.

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK

1

Zero too high

2

Amplifier output doesn't match test input

3

Both Failed