beautypg.com

Teledyne 3010PAC - Split architecture percent oxygen analyzer (EU - CENELEC) User Manual

Page 36

background image

4 Operation

Model 3010

PAC

4-2: Part I

Teledyne Analytical Instruments

4.2

Using the Data Entry and Function

Buttons

Data Entry Buttons: The < > arrow buttons select options from the

menu currently being displayed on the VFD screen. The selected option
blinks.

When the selected option includes a modifiable item, the

∆∆∆∆∆∇ arrow

buttons can be used to increment or decrement that modifiable item.

The

Enter

button is used to accept any new entries on the VFD screen.

The

Escape

button is used to abort any new entries on the VFD screen that

are not yet accepted by use of the

Enter

button.

Figure 4-1 shows the hierarchy of functions available to the operator via

the function buttons. The six function buttons on the analyzer are:

Analyze.

This is the normal operating mode. The analyzer

monitors the oxygen content of the sample, displays the
concentration of oxygen, and warns of any alarm conditions.

System.

The system function consists of six subfunctions that

regulate the internal operations of the analyzer:

LCD screen contrast

Auto-Cal setup

Password assignment

Self -Test initiation

Checking software version

Logging out.

Zero

. Used to set up a zero calibration.

Span.

Used to set up a span calibration.

Alarms.

Used to set the alarm setpoints and determine whether

each alarm will be active or defeated, HI or LO acting, latching,
and/or failsafe.

Range.

Used to set up three analysis ranges that can be switched

automatically with autoranging or used as individual fixed
ranges.

Any function can be selected at any time by pressing the appropriate

button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.

Contrast Function is DISABLED

(Refer to Section 1.3)