beautypg.com

Teledyne 3010PAC - Split architecture percent oxygen analyzer (EU - CENELEC) User Manual

Page 62

background image

2-2: Part II

2 Operational Theory

Model 3010PAC

Teledyne Analytical Instruments

being analyzed. The Micro-Fuel Cell is therefore a hybrid between a
battery and a true fuel cell. (All of the reactants are stored externally in a
true fuel cell.)

2.2.2 Anatomy of a Micro-Fuel Cell

A Micro-Fuel Cell (MFC) is a cylinder only 1¼ inches in diameter

and 1¼ inches thick. It is made of an extremely inert plastic, which can be
placed confidently in practically any environment or sample stream. The
cell is effectively sealed, although one end is permeable to oxygen in the
sample gas. The other end of the cell is a contact plate consisting of two
concentric foil rings. The rings mate with spring-loaded contacts in the
sensor block assembly and provide the electrical connection to the rest of
the analyzer. Figure 2-1 shows the external features of a typical cell.

Figure 2-1: Micro-Fuel Cell

Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which illus-

trates the following internal description.

Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale)