Teledyne 3010PAC - Split architecture percent oxygen analyzer (EU - CENELEC) User Manual

Page 42

background image

4 Operation

Model 3010

PAC

4-8: Part I

Teledyne Analytical Instruments

When you have finished typing the new password, press

Enter

. A

verification screen appears. The screen will prompt you to retype your
password for verification.

A A A A A
Retype PWD To Verify

Wait a moment. The entry screen will give you clearance to proceed.

A A A A A
TO Proceed

Use the arrow keys to retype your password and press

Enter

when

finished. Your password will be stored in the microprocessor and the system
will immediately switch to the

Analyze

screen, and you now have access to

all instrument functions.

If no alarms are tripped, the

Analyze

screen appears as:

0.0 % AnlZ
Range: 0 — 100

If an alarm is tripped, the second line will change to show which alarm

it is:

0.0 % Anlz
A L — 1

NOTE:If you previously logged off the system , you will now be

required to re-enter the password to gain access to Span,
Zero, Alarm, and Range functions.

4.3.4 Logout

The Logout function provides a convenient means of leaving the

analyzer in a password protected mode without having to shut the instrument
off. By entering Logout, you effectively log off the instrument leaving the
system protected against use until the password is reentered. To log out,
press the

System

button to enter the

System

function.

Contrast Auto—Cal
PSWD Logout More

Use the < > arrow keys to position the blinking over the Logout func-

tion, and press E

nter

to Log out. The screen will display the message:

Protected Until
Password Reentered

Contrast Function is DISABLED

(Refer to Section 1.3)