beautypg.com

Teledyne 3010PAC - Split architecture percent oxygen analyzer (EU - CENELEC) User Manual

Page 65

background image

Part II: 2-5

Oxygen Analyzers

Part II: Analysis Units

Teledyne Analytical Instruments

Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel Cell

2.2.6 Micro-Fuel Cell “Class”

TAI manufactures Micro-Fuel Cells with a variety of characteristics to

give the best possible performance for any given sample conditions. A few
typical Micro-Fuel Cells are listed below with their typical use and electri-
cal specifications.

2.2.6.1 Class A-3 Cell

The class A-3 cell is for use in applications where it is exposed con-

tinuously to carbon dioxide concentrations between 1 % and 100 % in the
sample gas.

Nominal output in air is 0.20 mA, and 90 % response time is 45 s.

Expected life in flue gas is 8 months.

2.2.6.2 Class A-5 Cell

The class A-5 cell is for use in applications where it is exposed inter-

mittently to carbon dioxide concentrations up to 100 % in the sample gas.

Nominal output in air is 0.19 mA, and 90 % response time is 45 s.

Expected life in flue gas is 8 months.