A-6

Appendix

Model 3000TB

­XL

Teledyne Analytical Instruments

BY-PASS:

To improve the system response, a by-pass can be added to increase the sample flow

rate to the analyzer by a factor of ten. A by-pass provides a sample flow path around the
analyzer of 2 - 18 SCFH. typically.

CALIBRATION GAS:

3000 series analyzer requirements for units with Auto-Cal options. The customer must

supply a control valves (or restrictors) for any SPAN or ZERO gas source which is attached
to the Auto-Cal ports. The valve should be adjusted to the same flow rate as the sample gas .
When restrictors are used, the gas pressure must be adjusted to achieve the proper flow rate.

OPERATION WITHOUT A RESTRICTOR DEVICE:

Operation without a restrictor device is not recommend as mentioned above. A

3000TB without any flow restrictor device was tested on 11-19-97. This results in a flow rate
of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the standard sample sys.

REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates at

various pressures.

TAI PART NUMBERS

RESTRICTOR KIT:

A68729

UNION (SS)

U11

LP. RESTRICTOR

R2323 ( LOW PRESSURE / VAC. SERVICE )

STD.. RESTRICTOR

R2324

BLUE DOT

NUT

N73

FERRULE

F73

FERRULE

F74

BOTH FERRULES ARE REQUIRED

CONVERSIONS:

1 PSI

= .04 INCHES OF MERCURY (in. Hg.)

1 SCFH

= 0.476 SLPM