3 Installation

Model 3000TB

­XL

3-10

Teledyne Analytical Instruments

The range in use (HI, MED, LO)

The span of the range (0-100 ppm, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

Each status output is followed by a carriage return and line feed.

Three input functions using RS-232 have been implemented to date.

They are described in Table 3-4.

Table 3-4: Commands via RS-232 Input

Command

Description

as

Immediately starts an autospan.

az

Immediately starts an autozero.

st

Toggling input. Stops/Starts any status message output from
the RS-232, until st is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-5 lists certain RS-232 values that are required by the 3000TB
implementation.

Table 3-5: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

3.3.9 Remote Sensor and Solenoid Valves

The 3000TB

­XL is a single-chassis instrument, which has its own sensor

and, in the –C option, its own gas-control solenoid valves. However, The
REMOTE SENSOR and SOLENOID RETURN connectors are provided
for use with a remote sensor and/or gas flow control valves, if desired. See
Figures 3-7 and 3-8.