3 Installation
Model 3000TB
XL
3-10
Teledyne Analytical Instruments
•
The range in use (HI, MED, LO)
•
The span of the range (0-100 ppm, etc)
•
Which alarms—if any—are disabled (AL–x DISABLED)
•
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 3-4.
Table 3-4: Commands via RS-232 Input
Command
Description
as
Immediately starts an autospan.
az
Immediately starts an autozero.
st
Toggling input. Stops/Starts any status message output from
the RS-232, until st
The RS-232 protocol allows some flexibility in its implementation.
Table 3-5 lists certain RS-232 values that are required by the 3000TB
implementation.
Table 3-5: Required RS-232 Options
Parameter
Setting
Baud
2400
Byte
8 bits
Parity
none
Stop Bits
1
Message Interval
2 seconds
3.3.9 Remote Sensor and Solenoid Valves
The 3000TB
XL is a single-chassis instrument, which has its own sensor
and, in the –C option, its own gas-control solenoid valves. However, The
REMOTE SENSOR and SOLENOID RETURN connectors are provided
for use with a remote sensor and/or gas flow control valves, if desired. See
Figures 3-7 and 3-8.