beautypg.com

Teledyne 3000TB-XL - Trace oxygen analyzer User Manual

Page 74

background image

A-2

Appendix

Model 3000TB

­XL

Teledyne Analytical Instruments

Accuracy:

±2% of full scale at constant temperature on all ranges except 0­1 ppm.

±5% of full scale over operating temperature range, except

0­10 ppm and 0­1 ppm range once thermal equilibrioum is reached.

±0.1 ppm on 0­1 ppm range once thermal equilibrium is reached.

Analog

Output: 0-1 V dc percent-of-range

0-1 V dc range ID.
4-20 mA dc percent-of-range, isolated
4-20 mA dc range ID, isolated

Password Access: Can be user-configured for password protection.

A-2 Recommended 2-Year Spare Parts List

Qty. Part Number

Description

1

C62959

Display PCB

1

D65295-A

Customer Interface PCB

1

C6

9367

Trace Preamplifier Board

1

C73870-B

Main PCB (4-20mADC Option)

3

F1295

Fuse, 4A, 250V, 5x20 mm, T (Slow Blow)

1

O38

O-ring

1

C69367

3000T

B-XL Amplifier

1

C6689-B2CXL

B2CXL Micro-Fuel Cell

(Std.)

1

B66378

B2C Adapter for Cell Block

Note: Orders for replacement parts should include the part number (if

available) and the model and serial number of the instrument for
which the parts are intended.