Trace Oxygen Analyzer

Installation 3

3-13

Teledyne Analytical Instruments

2. Hold the fitting body steady with a backup wrench, and with

another wrench rotate the nut another 1

1

/

4

turns.

Figure 3-10: Gas Connector Panel

SAMPLE IN: In the standard model, gas connections are made at the

SAMPLE IN

and

EXHAUST OUT

connections. Calibration gases must be

Tee'd into the Sample inlet with appropriate valves.

The gas pressure in should be reasonably regulated. Pressures between

3 and 40 psig are acceptable as long as the pressure, once established, will
keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4
SLPM). Exact figures will depend on your process.

If greater flow is required for improved response time, install a bypass

in the sampling system upstream of the analyzer input.

Note: If the unit is for vacuum service, the above numbers apply

instead to the vacuum at the EXHAUST OUT connector, de-
scribed below, with minus signs before the pressure readings.

ZERO IN and SPAN IN: These are additional ports for inputting span

gas and zero gas. There are electrically operated valves inside for automatic
switching between sample and calibration gases. These valves are com-
pletely under control of the 3000TB

­XL Electronics. They can be externally

controlled only indirectly through the Remote Cal Inputs, described below.

Pressure, flow, and safety considerations are the same as prescribed for

the SAMPLE IN inlet, above.