Teledyne 3000TB-XL - Trace oxygen analyzer User Manual

Page 10

background image

1-2

1 Introduction

Model 3000TB

­XL

Teledyne Analytical Instruments

316 Stainless steel cell block and sample system—all wetted
parts.

Advanced Micro-Fuel Cell for trace analysis. Standard

Insta Trace cell

has rapid recovery from air to low ppm during initial installation.

Versatile analysis over a wide range of applications.

Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.

Three user definable output ranges (from 0-1 ppm through 0-
25 % allow best match to users process and equipment.

Air-calibration range for convenient spanning at 20.9 %.

Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.

Two adjustable concentration alarms and a system failure alarm.

Extensive self-diagnostic testing, at startup and on demand, with
continuous power-supply monitoring.

RS-232 serial digital port for use with a computer or other digital
communication device.

Analog outputs for concentration and range identification.
(0-1 V dc standard, and isolated 4–20 mA dc optional.)

1.4

Model Designations

3000TB

­XL: Standard model.

3000TB

­XL-C: In addition to all standard features, this model also has

separate ports for zero and span gases, and built-in control

valves. The internal valves are entirely under the control of

the 3000TB­XL electronics, to automatically switch between

gases in synchronization with the analyzer’s operations.