Appendix

A-1 Specifications

Packaging: NEMA-4, weather-proof, bulkhead mount.

Sensor:

B2­CXL (C6689­B2CXL) Micro-Fuel Cell (std.)

Other cells optional.

Cell Block: 316 stainless steel.

Ranges: Three user definable ranges from

0–1ppm to 0–250,000ppm,

plus air calibration range of 0-250,000 ppm (25 %).

Autoranging with range ID output.

Sample System: Flow indicator visible from front of unit.

Positive pressure service.

Vacuum service (optional).

Auto Cal / Auto Zero. (Available with optional,
electrically operated valves

except XL option.)

Alarms: One system-failure alarm contact to detect

power failure.

Two adjustable concentration threshold alarms
with fully programmable setpoints.

Diagnostics: Start-up or on-demand, comprehensive, self

testing function initiated by keyboard or
remote command.

Displays: 2 line by 20 alphanumeric, VFD screen, and

one 5 digit LED display.

Digital Interface: Full duplex RS-232 communications port.

Power: 110 or 220 VAC at 50 or 60 Hz.

Operating Temperature:

5-35 °C