beautypg.com

3 the system function, Trace oxygen analyzer operation 4, Teledyne analytical instruments – Teledyne 3000TB-XL - Trace oxygen analyzer User Manual

Page 43: Analyze system span zero alarms rang e

background image

Trace Oxygen Analyzer

Operation 4

4-3

Teledyne Analytical Instruments

Figure 4-1: Hierarchy of Functions and Subfunctions

4.3

The System Function

The subfuctions of the

System

function are described below. Specific

procedures for their use follow the descriptions:

Auto-Cal: Used to define an automatic calibration sequence
and/or start an Auto-Cal.

PSWD: Security can be established by choosing a 5 digit
password (PSWD) from the standard ASCII character set. (See
Installing or Changing a Password, below, for a table of ASCII
characters available.) Once a unique password is assigned and
activated, the operator MUST enter the UNIQUE password to
gain access to set-up functions which alter the instrument's
operation, such as setting the instrument span or zero setting,
adjusting the alarm setpoints, or defining analysis ranges.

Set Instrument

Span

Set Instrument

Zero

Set Alarm

Setpoints

Confrigure Mode

of Alarm

Operation

Define Analysis

Ranges

Perform Oxygen

Analysis of

the Sample

ANALYZE

SYSTEM

SPAN

ZERO

ALARMS

RANG

E

TRAK/HLD

Perform

Self-Diagnostic

Test

Initiate

Automatic

Calibration

Set Password

Logout

Show Negative