beautypg.com

Teledyne 3000TB-XL - Trace oxygen analyzer User Manual

Page 34

background image

3 Installation

Model 3000TB

­XL

3-8

Teledyne Analytical Instruments

ZERO:

Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the

Zero

mode. Either side may be

grounded at the source of the signal. 0 to 1 volt across the
terminals allows

Zero

mode to terminate when done. A

synchronous signal must open and close the external zero
valve appropriately. See section 3.3.9 Remote Sensor and
Solenoid Valves
. (With the –C option, the internal valves
automatically operate synchronously.)

SPAN:

Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the

Span

mode. Either side may be

grounded at the source of the signal. 0 to 1 volt across the
terminals allows

Span

mode to terminate when done. A

synchronous signal must open and close the external span
valve appropriately. See 3.3.9 Remote Sensor and Solenoid
Valves
. (With the –C option, the internal valves automatically
operate synchronously.)

Cal Contact: This relay contact is closed while analyzer is spanning

and/or zeroing. (See Remote Calibration Protocol below.)

Remote Calibration Protocol: To properly time the Digital Remote

Cal Inputs to the Model 3000TB

­XL Analyzer, the customer's controller must

monitor the CAL CONTACT relay.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal

Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It

will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact

closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.

For example:

1) Test the CRC. When the CRC is open, Send a zero command

until the CRC closes (The CRC will quickly close.)

2) When the CRC closes, remove the zero command.

3) When CRC opens again, send a span command until the CRC

closes. (The CRC will quickly close.)

4) When the CRC closes, remove the span command.

When CRC opens again, zero and span are done, and the sample is

being analyzed.