BY-PASS:
To improve the system response, a by-pass can be added to increase the sample
flow rate to the analyzer by a factor of ten. A by-pass provides a sample flow path around
the analyzer of 2 - 18 SCFH. typically.
CALIBRATION GAS:
3000 series analyzer requirements for units with Auto-Cal options. The customer
must supply a control valves (or restrictors) for any SPAN or ZERO gas source which is
attached to the Auto-Cal ports. The valve should be adjusted to the same flow rate as the
sample gas . When restrictors are used, the gas pressure must be adjusted to achieve the
proper flow rate.
OPERATION WITHOUT A RESTRICTOR DEVICE:
Operation without a restrictor device is not recommend as mentioned above. A
3000PB without any flow restrictor device was tested on 11-19-97. This results in a flow
rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the standard sample sys.
REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates at
various pressures.
TAI PART NUMBERS
RESTRICTOR KIT:
A68729
UNION (SS)
U11
LP. RESTRICTOR
R2323 ( LOW PRESSURE / VAC. SERVICE )
STD.. RESTRICTOR
R2324
BLUE DOT
NUT
N73
FERRULE
F73
FERRULE
F74
BOTH FERRULES ARE
REQUIRED
CONVERSIONS:
1 PSI
=
2.04 INCHES OF MERCURY (in. Hg.)
1 SCFH =
0.476 SLPM